<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
  <dc:creator>Lang PZ</dc:creator>
  <dc:creator>Thulasi P</dc:creator>
  <dc:creator>Khandelwal SS</dc:creator>
  <dc:creator>Hafezi F</dc:creator>
  <dc:creator>Randleman JB</dc:creator>
  <dc:date>2018</dc:date>
  <dc:description xmlns:ns0="xml" ns0:lang="en">PURPOSE
To evaluate the correlation between anterior axial curvature difference maps following corneal cross-linking (CXL) for progressive keratoconus obtained from Scheimpflug-based tomography and Placido-based topography.


DESIGN
Between-devices reliability analysis of randomized clinical trial data.


METHODS
Corneal imaging was collected at a single-center institution preoperatively and at 3, 6, and 12 months postoperatively using Scheimpflug-based tomography (Pentacam; Oculus Inc, Lynnwood, Washington, USA) and scanning-slit, Placido-based topography (Orbscan II; Bausch &amp; Lomb, Rochester, New York, USA) in patients with progressive keratoconus receiving standard protocol CXL (3 mW/cm2 for 30 minutes). Regularization index (RI), absolute maximum keratometry (K Max), and change in K Max (ΔK Max) were compared between the 2 devices at each time point.


RESULTS
Fifty-one eyes from 36 patients were evaluated at all time points. Values were significantly different at all time points (56.01 ± 5.3 diopters [D] Scheimpflug vs 55.04 ± 5.1 D scanning-slit preoperatively [P = .003]; 54.58 ± 5.3 D Scheimpflug vs 53.12 ± 4.9 D scanning-slit at 12 months [P &lt; .0001]) but strongly correlated between devices (r = 0.90-0.93) at all time points. The devices were not significantly different at any time point for either ΔK Max or RI but were poorly correlated at all time points (r = 0.41-0.53 for ΔK Max, r = 0.29-0.48 for RI). At 12 months, 95% limits of agreement were 7.51 D for absolute K Max, 8.61 D for ΔK Max, and 19.86 D for RI.


CONCLUSIONS
Measurements using Scheimpflug and scanning-slit Placido-based technology are correlated but not interchangeable. Both devices appear reasonable for separately monitoring the cornea's response to CXL; however, caution should be used when comparing results obtained with one measuring technology to the other.</dc:description>
  <dc:identifier>https://sonar.ch/global/documents/232710</dc:identifier>
  <dc:language>eng</dc:language>
  <dc:relation>info:eu-repo/semantics/altIdentifier/doi/10.1016/j.ajo.2018.04.018</dc:relation>
  <dc:relation>info:eu-repo/semantics/altIdentifier/pmid/29729256</dc:relation>
  <dc:source>American journal of ophthalmology. - 2018</dc:source>
  <dc:subject xmlns:ns1="xml" ns1:lang="en">Adult</dc:subject>
  <dc:subject xmlns:ns2="xml" ns2:lang="en">Cornea</dc:subject>
  <dc:subject xmlns:ns3="xml" ns3:lang="en">Corneal Pachymetry</dc:subject>
  <dc:subject xmlns:ns4="xml" ns4:lang="en">Corneal Topography</dc:subject>
  <dc:subject xmlns:ns5="xml" ns5:lang="en">Cross-Linking Reagents</dc:subject>
  <dc:subject xmlns:ns6="xml" ns6:lang="en">Equipment Design</dc:subject>
  <dc:subject xmlns:ns7="xml" ns7:lang="en">Female</dc:subject>
  <dc:subject xmlns:ns8="xml" ns8:lang="en">Follow-Up Studies</dc:subject>
  <dc:subject xmlns:ns9="xml" ns9:lang="en">Humans</dc:subject>
  <dc:subject xmlns:ns10="xml" ns10:lang="en">Imaging, Three-Dimensional</dc:subject>
  <dc:subject xmlns:ns11="xml" ns11:lang="en">Keratoconus</dc:subject>
  <dc:subject xmlns:ns12="xml" ns12:lang="en">Male</dc:subject>
  <dc:subject xmlns:ns13="xml" ns13:lang="en">Photochemotherapy</dc:subject>
  <dc:subject xmlns:ns14="xml" ns14:lang="en">Prospective Studies</dc:subject>
  <dc:subject xmlns:ns15="xml" ns15:lang="en">Reproducibility of Results</dc:subject>
  <dc:title xmlns:ns16="xml" ns16:lang="en">Comparing Change in Anterior Curvature After Corneal Cross-linking Using Scanning-slit and Scheimpflug Technology.</dc:title>
  <dc:type>http://purl.org/coar/resource_type/c_6501</dc:type>