  <dc:description xmlns:ns0="xml" ns0:lang="en">PURPOSE
We present a prototype of the hand-held dynamic contour tonometer (HH-DCT) and prospectively compare this HH-DCT with the well-established Perkins applanation tonometer (PAT) and the TonoPenXL (TPXL).


METHODS
In a prospective, single-centre, randomized study, intraocular pressure (IOP) readings were taken in random order using HH-DCT, PAT and TPXL tonometers. Intra-observer variability was calculated for each observer and compared between three experienced ophthalmologists and an inexperienced medical student.


RESULTS
Ninety-two corneas of 92 healthy participants were enrolled. IOP [mean mmHg ± standard deviation (SD)] as measured by HH-DCT was 16.97 ± 2.71, by PAT 13.98 ± 2.52 and by TPXL 13.34 ± 2.68. The range of three consecutive IOP readings differed significantly between the devices [p &lt; 0.001; mean range: 1.45 ± 1.07 (HH-DCT), 1.87 ± 0.97 (PAT) and 2.08 ± 1.77 (TPXL)]. There was no difference of the range in all devices between the ophthalmologists and the medical student (HH-DCT p = 0.68, PAT p = 0.54, TPXL p = 0.48).


CONCLUSION
IOP readings measured by HH-DCT are significantly higher than by PAT and TPXL. The differences of IOP measurements are in good accordance with previous studies using the slit-lamp-mounted DCT (SL-DCT) and Goldmann Applanation Tonometry, where SL-DCT readings were 1-3.2 mmHg higher. HH-DCT seems to give more constant results, which can be seen in the lower intra-observer variability compared to PAT and TPXL.</dc:description>
  <dc:format>application/pdf</dc:format>
  <dc:identifier>https://sonar.ch/global/documents/236037</dc:identifier>
  <dc:language>eng</dc:language>
  <dc:relation>info:eu-repo/semantics/altIdentifier/doi/10.1111/j.1755-3768.2009.01625.x</dc:relation>
  <dc:relation>info:eu-repo/semantics/altIdentifier/pmid/21348963</dc:relation>
  <dc:rights>info:eu-repo/semantics/openAccess</dc:rights>
  <dc:source>Acta ophthalmologica. - 2011</dc:source>
  <dc:subject xmlns:ns1="xml" ns1:lang="en">Adult</dc:subject>
  <dc:subject xmlns:ns2="xml" ns2:lang="en">Female</dc:subject>
  <dc:subject xmlns:ns3="xml" ns3:lang="en">Humans</dc:subject>
  <dc:subject xmlns:ns4="xml" ns4:lang="en">Intraocular Pressure</dc:subject>
  <dc:subject xmlns:ns5="xml" ns5:lang="en">Male</dc:subject>
  <dc:subject xmlns:ns6="xml" ns6:lang="en">Middle Aged</dc:subject>
  <dc:subject xmlns:ns7="xml" ns7:lang="en">Observer Variation</dc:subject>
  <dc:subject xmlns:ns8="xml" ns8:lang="en">Prospective Studies</dc:subject>
  <dc:subject xmlns:ns9="xml" ns9:lang="en">Reproducibility of Results</dc:subject>
  <dc:subject xmlns:ns10="xml" ns10:lang="en">Tonometry, Ocular</dc:subject>
  <dc:subject xmlns:ns11="xml" ns11:lang="en">Young Adult</dc:subject>
  <dc:title xmlns:ns12="xml" ns12:lang="en">Hand-held dynamic contour tonometry.</dc:title>