  <dc:creator>Dobrev I</dc:creator>
  <dc:creator>Sim JH</dc:creator>
  <dc:creator>Pfiffner F</dc:creator>
  <dc:creator>Huber AM</dc:creator>
  <dc:creator>Röösli C</dc:creator>
  <dc:date>2018</dc:date>
  <dc:description xmlns:ns0="xml" ns0:lang="en">OBJECTIVES
Evaluation of the transfer function efficiency of a newly-developed piezo-electric actuator for active subcutaneous bone conduction hearing aid.


METHODS
The experiments were conducted on four Thiel embalmed whole head cadaver specimens. A novel actuator based on piezo-electric transduction (PZTA), part of a subcutaneous bone conduction hearing aid device, was sequentially implanted on three locations: 1) Immediately posterior to pinna; 2) 50-60 mm posterior to pinna, approximately the same distance as between the BAHA (bone anchored hearing aid) location and the ear canal, but the same horizontal level as location 1; 3) the traditional BAHA location. Using a single point 3-dimensional laser Doppler vibrometer (LDV) system, three types of motion measurements were performed at the cochlear promontory for each stimulation location: 1) ipsilateral side, 2) contralateral side, 3) measurements 1 and 2 were repeated after mastoidectomy on the ipsilateral side.


RESULTS
On average, stimulation at locations 1 and 2 show a trend for higher promontory motion relative to location 3 (BAHA location) above 1 kHz. Stimulation at location 1 had an average improvement of 1-6 dB at 2-4 kHz, and 1-18 dB at 6-8 kHz. The spatial composition of the motion showed significant contributions from both in-plane and out-of-plane (along ear canal) motion components, with in-plane components being dominant at mid and high frequencies for locations 2 and 3. Stimulation at locations 1 and 3 produced similar transcranial attenuation at mid frequencies (0.6-4 kHz), with a potential trend of higher attenuation (seen in 3 or the 4 samples) for location 1 at higher frequencies (&gt;4 kHz). The mastoidectomy affected negatively mostly the high frequencies (6-8 kHz) for stimulation at location 1, with no significant change for location 3.


CONCLUSION
The sound transfer function efficacy of a novel subcutaneous bone conduction device has been quantified, and the influence of stimulation location and mastoidectomy have been analyzed based on promontory motion in Thiel-preserved cadaver heads.</dc:description>
  <dc:format>application/pdf</dc:format>
  <dc:identifier>https://sonar.ch/global/documents/88361</dc:identifier>
  <dc:language>eng</dc:language>
  <dc:relation>info:eu-repo/semantics/altIdentifier/doi/10.1016/j.heares.2018.10.003</dc:relation>
  <dc:relation>info:eu-repo/semantics/altIdentifier/pmid/30343248</dc:relation>
  <dc:rights>info:eu-repo/semantics/openAccess</dc:rights>
  <dc:source>Hearing research. - 2018</dc:source>
  <dc:subject xmlns:ns1="xml" ns1:lang="en">Bone conduction</dc:subject>
  <dc:subject xmlns:ns2="xml" ns2:lang="en">Bone conduction hearing aid</dc:subject>
  <dc:subject xmlns:ns3="xml" ns3:lang="en">Cadaver head</dc:subject>
  <dc:subject xmlns:ns4="xml" ns4:lang="en">Laser Doppler vibrometer</dc:subject>
  <dc:subject xmlns:ns5="xml" ns5:lang="en">Mastoidectomy</dc:subject>
  <dc:subject xmlns:ns6="xml" ns6:lang="en">Piezoelectric actuator</dc:subject>
  <dc:subject xmlns:ns7="xml" ns7:lang="en">Transcranial attenuation</dc:subject>
  <dc:subject xmlns:ns8="xml" ns8:lang="en">Acoustic Stimulation</dc:subject>
  <dc:subject xmlns:ns9="xml" ns9:lang="en">Bone Conduction</dc:subject>
  <dc:subject xmlns:ns10="xml" ns10:lang="en">Bone-Anchored Prosthesis</dc:subject>
  <dc:subject xmlns:ns11="xml" ns11:lang="en">Cadaver</dc:subject>
  <dc:subject xmlns:ns12="xml" ns12:lang="en">Hearing Aids</dc:subject>
  <dc:subject xmlns:ns13="xml" ns13:lang="en">Humans</dc:subject>
  <dc:subject xmlns:ns14="xml" ns14:lang="en">Mastoidectomy</dc:subject>
  <dc:subject xmlns:ns15="xml" ns15:lang="en">Materials Testing</dc:subject>
  <dc:subject xmlns:ns16="xml" ns16:lang="en">Motion</dc:subject>
  <dc:subject xmlns:ns17="xml" ns17:lang="en">Prosthesis Design</dc:subject>
  <dc:subject xmlns:ns18="xml" ns18:lang="en">Skull</dc:subject>
  <dc:subject xmlns:ns19="xml" ns19:lang="en">Sound</dc:subject>
  <dc:subject xmlns:ns20="xml" ns20:lang="en">Vibration</dc:subject>
  <dc:title xmlns:ns21="xml" ns21:lang="en">Performance evaluation of a novel piezoelectric subcutaneous bone conduction device.</dc:title>