Comparison of imaging methods used for dental implant osseous integration assessment.
Journal article

Comparison of imaging methods used for dental implant osseous integration assessment.

Show more…
  • 2013-07-11
Published in:
  • Journal of materials science. Materials in medicine. - 2013
English Two different imaging techniques used to determine bone tissue response to dental implants were compared. Dental implants were implanted into the maxillae of 18 pigs, which were sacrificed after 4, 8 and 12 weeks. Implants with surrounding bone tissue were retrieved for methyl methacrylate histology and contact radiography. On identical sections peri-implant bone density and bone implant contact (BIC) ratio were assessed with two different imaging methods. Evaluation of Giemsa eosin stained and contact radiographed sections showed direct osseous integration for all implants and both methods showed a strong correlation with correlation coefficient r = 0.930 (P < 0.0001) for peri-implant bone density and r = 0.817 (P < 0.0001) for bone implant contact ratio. While the two imaging methods showed moderate differences for peri-implant bone density there were significant differences between the BIC values determined. In general, contact radiography tends to underestimate BIC for approximately 4.5 % (P = 0.00003).
Language
  • English
Open access status
closed
Identifiers
Persistent URL
https://sonar.ch/global/documents/10748
Statistics

Document views: 14 File downloads: