Evaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus).

  • Steinmetz HW Division of Zoo Animals and Exotic Pets, Vetsuisse Faculty, University of Zürich, Winterthurerstrasse 260, 8057 Zürich, Switzerland. hsteinmetz@vetclinics.unizh.ch
  • Vogt R
  • Kästner S
  • Riond B
  • Hatt JM
Show more…
  • 2007-07-05
Published in:
  • Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc. - 2007
English The i-STAT portable clinical analyzer (PCA) was evaluated for performance in avian species. With the EG7+ cartridge, which provided results for hydrogen ion concentration, oxygen tension, carbon dioxide tension, sodium, potassium, ionized calcium, hematocrit, and various calculated parameters, analytical accuracy and precision were tested by comparing obtained values to those of established traditional blood gas and chemistry analyzers. Deming's regression and bias plots were used to compare i-STAT results with those obtained by laboratory professionals using benchtop analyzers. The reliability of the i-STAT PCA with EG7+ cartridges was good, with 0-5.7% system failures in measured values. Regression statistics were good for all blood gas analytes and acceptable for electrolytes and calculated parameters, except for potassium and base excess, for which the regression data or the discrepancy between the methods was too large. The system was reliable and easy to use and had an overall acceptable accuracy in avian species. These features, together with portability and small required blood volumes, make the i-STAT suitable for point-of-care use in critical avian patients, although single values require careful interpretation.
Language
  • English
Open access status
bronze
Identifiers
Persistent URL
https://sonar.ch/global/documents/235646
Statistics

Document views: 15 File downloads:
  • fulltext.pdf: 0