The text has been copied to the clipboard
Journal article

Combining phase images from array coils using a short echo time reference scan (COMPOSER).

  • Robinson SD High Field Magnetic Resonance Centre, Medical University of Vienna, Austria.
  • Dymerska B High Field Magnetic Resonance Centre, Medical University of Vienna, Austria.
  • Bogner W High Field Magnetic Resonance Centre, Medical University of Vienna, Austria.
  • Barth M The University of Queensland, Centre for Advanced Imaging, Brisbane, Australia.
  • Zaric O High Field Magnetic Resonance Centre, Medical University of Vienna, Austria.
  • Goluch S High Field Magnetic Resonance Centre, Medical University of Vienna, Austria.
  • Grabner G High Field Magnetic Resonance Centre, Medical University of Vienna, Austria.
  • Deligianni X Department of Radiology, Division of Radiological Physics, University of Basel Hospital, Basel, Switzerland.
  • Bieri O Department of Radiology, Division of Radiological Physics, University of Basel Hospital, Basel, Switzerland.
  • Trattnig S High Field Magnetic Resonance Centre, Medical University of Vienna, Austria.
Show more…
  • 2015-12-30
Published in:
  • Magnetic resonance in medicine. - 2017
English PURPOSE
To develop a simple method for combining phase images from multichannel coils that does not require a reference coil and does not entail phase unwrapping, fitting or iterative procedures.


THEORY AND METHODS
At very short echo time, the phase measured with each coil of an array approximates to the phase offset to which the image from that coil is subject. Subtracting this information from the phase of the scan of interest matches the phases from the coils, allowing them to be combined. The effectiveness of this approach is quantified in the brain, calf and breast with coils of diverse designs.


RESULTS
The quality of phase matching between coil elements was close to 100% with all coils assessed even in regions of low signal. This method of phase combination was similar in effectiveness to the Roemer method (which needs a reference coil) and was superior to the rival reference-coil-free approaches tested.


CONCLUSION
The proposed approach-COMbining Phase data using a Short Echo-time Reference scan (COMPOSER)-is a simple and effective approach to reconstructing phase images from multichannel coils. It requires little additional scan time, is compatible with parallel imaging and is applicable to all coils, independent of configuration. Magn Reson Med 77:318-327, 2017. © 2015 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.