Journal article

Improving the quality of XAFS data.

  • Abe H High Energy Accelerator Research Organization (KEK), Tsukuba, Japan.
  • Aquilanti G Elettra-Sincrotrone Trieste, Basovizza, Italy.
  • Boada R Diamond Light Source, Oxfordshire, UK.
  • Bunker B University of Notre Dame, Notre Dame, IN, USA.
  • Glatzel P European Synchrotron Radiation Facility, Grenoble, France.
  • Nachtegaal M Paul Scherrer Institute, Villigen, Switzerland.
  • Pascarelli S European Synchrotron Radiation Facility, Grenoble, France.
Show more…
  • 2018-07-07
Published in:
  • Journal of synchrotron radiation. - 2018
English Following the Q2XAFS Workshop and Satellite to IUCr Congress 2017 on `Data Acquisition, Treatment, Storage - quality assurance in XAFS spectroscopy', a summary is given of the discussion on different aspects of a XAFS experiment that affect data quality. Some pertinent problems ranging from sources and minimization of noise to harmonic contamination and uncompensated monochromator glitches were addressed. Also, an overview is given of the major limitations and pitfalls of a selection of related methods, such as photon-out spectroscopies and energy-dispersive XAFS, and of increasingly common applications, namely studies at high pressure, and time-resolved investigations of catalysts in operando. Advice on how to avoid or deal with these problems and a few good practice recommendations are reported, including how to correctly report results.
Language
  • English
Open access status
hybrid
Identifiers
Persistent URL
https://sonar.ch/global/documents/77392
Statistics

Document views: 49 File downloads: