Magnetic Field Stabilization for Magnetically Shielded Volumes by External Field Coils.
-
Brys T
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Czekaj S
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Daum M
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Fierlinger P
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
George D
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Henneck R
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Hochman Z
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Kasprzak M
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Kohlik K
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Kirch K
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Kuzniak M
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Kuehne G
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Pichlmaier A
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Siodmok A
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Szelc A
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
-
Tanner L
Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland.
Show more…
Published in:
- Journal of research of the National Institute of Standards and Technology. - 2005
English
For highly sensitive magnetic measurements, e.g., a measurement of the neutron electric dipole moment (EDM), the magnetic field has to be stable in time on a level below picoTesla. One of several measures we employ to achieve this uses an external field coil system which can stabilize the ambient external field at a predefined value. Here we report on the construction and characterization of such a system in the magnetic test facility at PSI. The system actively stabilizes the field along the axis of the EDM experiment by means of four coils in a Helmholtz-like configuration. Additional coils serve to compensate for transverse ambient field components. Because of the long integration times in the EDM experiment (about 100 s or more) only slow disturbances have to be corrected for. The performance of the system has been measured using static and moving magnetic sources and suppression factors in excess of 200 have been observed.
-