Optimal cut-off criteria for duplex ultrasound compared with computed tomography angiography for the diagnosis of restenosis in stented carotid arteries in the international carotid stenting study.
-
Bosch FT
Department of Neurology, Academic Medical Center, Amsterdam, the Netherlands.
-
Hendrikse J
Department of Neurology and Neurosurgery, UMCU, Utrecht, the Netherlands.
-
Davagnanam I
UCL Institute of Neurology, The National Hospital for Neurology and Neurosurgery, London, UK.
-
Bonati LH
Department of Neurology, University Hospital Basel, Basel, Switzerland.
-
van der Lugt A
Department of Radiology, Erasmus MC, Rotterdam, the Netherlands.
-
van der Worp HB
Department of Neurology and Neurosurgery, UMCU, Utrecht, the Netherlands.
-
de Borst GJ
Department of Neurology and Neurosurgery, UMCU, Utrecht, the Netherlands.
-
Mali W
Department of Neurology and Neurosurgery, UMCU, Utrecht, the Netherlands.
-
Brown MM
UCL Institute of Neurology, The National Hospital for Neurology and Neurosurgery, London, UK.
-
Nederkoorn PJ
Department of Neurology, Academic Medical Center, Amsterdam, the Netherlands.
Show more…
Published in:
- European stroke journal. - 2017
English
Introduction
Previous studies that reported duplex-ultrasound cut-off criteria, based on blood velocity parameters, for the degree of stenosis in a stented carotid artery were either retrospective, or the reference test was carried out only when a patient was suspected of having restenosis at duplex ultrasound, which is likely to have resulted in verification bias. We performed a prospective study of diagnostic accuracy to find new blood velocity cut-offs in duplex ultrasound for in-stent restenosis.
Patients and methods
Stented patients within the international carotid stenting study were eligible. Patients had a carotid computed tomography angiography in addition to routine duplex ultrasound performed at a yearly follow-up. Duplex-ultrasound bloodflow velocity parameters were compared to the degree of stenosis on computed tomography angiography. The results were analysed using receiver-operating-characteristic curves.
Results
We included 103 patients in this study. On computed tomography angiography, 30 (29.1%) patients had a 30%-49% in-stent restenosis, 21 (20.4%) patients had 50%-69% in-stent restenosis and 5 (4.9%) patients a ≥70% in-stent restenosis. The cut-off value ≥50% stenosis was a peak systolic velocity of 125 cm/s (sensitivity: 63% (95% CI: 41-79), specificity: 83% (95% CI: 72-90)).
Discussion
This study provides a level 2b evidence for new cut-off values for in-stent restenosis. Unfortunately, we could not say anything about severe stenosis because of the low number of severe stenosis after one year.
Conclusions
The 125 cm/s cut-off value on duplex ultrasound is lower than found in previous studies and equal to unstented arteries. Duplex-ultrasound measurements made in stented carotid arteries should not be corrected for the presence of a stent when determining the degree of stenosis.
-