Comparing Change in Anterior Curvature After Corneal Cross-linking Using Scanning-slit and Scheimpflug Technology.
Journal article

Comparing Change in Anterior Curvature After Corneal Cross-linking Using Scanning-slit and Scheimpflug Technology.

  • Lang PZ USC Roski Eye Institute, Keck School of Medicine of the University of Southern California, Los Angeles, California, USA.
  • Thulasi P Emory Eye Center, Atlanta, Georgia, USA.
  • Khandelwal SS Baylor College of Medicine, Cullen Eye Institute, Houston, Texas, USA.
  • Hafezi F USC Roski Eye Institute, Keck School of Medicine of the University of Southern California, Los Angeles, California, USA; Faculty of Medicine, University of Geneva, Geneva, Switzerland; ELZA Institute, Dietikon/Zurich, Switzerland; Department of Ophthalmology, Wenzhou Medical University, Wenzhou, China.
  • Randleman JB USC Roski Eye Institute, Keck School of Medicine of the University of Southern California, Los Angeles, California, USA. Electronic address: randlema@usc.edu.
  • 2018-05-06
Published in:
  • American journal of ophthalmology. - 2018
English PURPOSE
To evaluate the correlation between anterior axial curvature difference maps following corneal cross-linking (CXL) for progressive keratoconus obtained from Scheimpflug-based tomography and Placido-based topography.


DESIGN
Between-devices reliability analysis of randomized clinical trial data.


METHODS
Corneal imaging was collected at a single-center institution preoperatively and at 3, 6, and 12 months postoperatively using Scheimpflug-based tomography (Pentacam; Oculus Inc, Lynnwood, Washington, USA) and scanning-slit, Placido-based topography (Orbscan II; Bausch & Lomb, Rochester, New York, USA) in patients with progressive keratoconus receiving standard protocol CXL (3 mW/cm2 for 30 minutes). Regularization index (RI), absolute maximum keratometry (K Max), and change in K Max (ΔK Max) were compared between the 2 devices at each time point.


RESULTS
Fifty-one eyes from 36 patients were evaluated at all time points. Values were significantly different at all time points (56.01 ± 5.3 diopters [D] Scheimpflug vs 55.04 ± 5.1 D scanning-slit preoperatively [P = .003]; 54.58 ± 5.3 D Scheimpflug vs 53.12 ± 4.9 D scanning-slit at 12 months [P < .0001]) but strongly correlated between devices (r = 0.90-0.93) at all time points. The devices were not significantly different at any time point for either ΔK Max or RI but were poorly correlated at all time points (r = 0.41-0.53 for ΔK Max, r = 0.29-0.48 for RI). At 12 months, 95% limits of agreement were 7.51 D for absolute K Max, 8.61 D for ΔK Max, and 19.86 D for RI.


CONCLUSIONS
Measurements using Scheimpflug and scanning-slit Placido-based technology are correlated but not interchangeable. Both devices appear reasonable for separately monitoring the cornea's response to CXL; however, caution should be used when comparing results obtained with one measuring technology to the other.