The text has been copied to the clipboard
Accuracy of Computer-Guided Template-Based Implant Surgery: A Computed Tomography-Based Clinical Follow-Up Study.
Journal article

Accuracy of Computer-Guided Template-Based Implant Surgery: A Computed Tomography-Based Clinical Follow-Up Study.

  • Schelbert T Doctoral Student, Clinic of Cranio-Maxillo-Facial and Oral Surgery, University Hospital Zurich, Zurich, Switzerland.
  • Gander T Senior Physician, Clinic of Cranio-Maxillo-Facial and Oral Surgery, University Hospital Zurich, Zurich, Switzerland.
  • Blumer M Senior Physician, Clinic of Cranio-Maxillo-Facial and Oral Surgery, University Hospital Zurich, Zurich, Switzerland.
  • Jung R Head of Division of Implantology and Vice Chairman of Clinic of Fixed and Removable Prosthodontics and Dental Material Science, Center of Dental Medicine, University of Zurich, Zurich, Switzerland.
  • Rücker M Professor and Head of Clinic of Cranio-Maxillo-Facial and Oral Surgery, University Hospital Zurich, Zurich, Switzerland.
  • Rostetter C Assistant Physician, Clinic of Cranio-Maxillo-Facial and Oral Surgery, University Hospital Zurich, Zurich, Switzerland.
Show more…
  • 2019-09-14
Published in:
  • Implant dentistry. - 2019
English OBJECTIVE
The aim of this clinical study was to analyze the accuracy of computer-guided implant surgery.


MATERIALS AND METHODS
Assisted by computed tomography (CT)-based planning software and navigational templates, 16 patients successfully received 26 dental implants. Each implant parameter (a-d) was calculated based on superimposed preoperative and postoperative cone beam CT scans: (a) deviation at entry point; (b) deviation at apex; (c) angular deviation; and (d) depth deviation.


RESULTS
Mean central deviation at implant entry point and apex was 0.91 mm (standard error [SE] = 0.11 mm; 95% confidence interval [CI]: 0.69-1.13) and 1.22 mm (SE = 0.11 mm; 95% CI: 0.99-1.45), respectively. Mean angulation deviation was 4.11 degrees (SE = 0.52 degrees; 95% CI: 3.04-5.17) and the average depth deviation was 0.65 mm (SE = 0.11 mm; 95% CI: 0.42-0.87). For the total number of implants placed, the maximum error was 2.34 mm at entry point, 2.71 mm at apex, 9.44 degrees in angular deviation, and 2.00 mm in depth deviation.


CONCLUSION
Great accuracy was reached even in advanced cases with prior bone augmentation and complex traumas. This leads to the conclusion that particularly in advanced cases, computer-guided implantation can be beneficial.