Prevention of cement leakage into the hip joint by a standard cement plug during PFN-A cement augmentation: a technical note.
Journal article

Prevention of cement leakage into the hip joint by a standard cement plug during PFN-A cement augmentation: a technical note.

  • Hanke M Department of Orthopaedic and Trauma Surgery, University of Bern, Inselspital, Freiburgstrasse 3, 3010, Bern, Switzerland.
  • Djonov V Institute of Anatomy, University of Bern, Bern, Switzerland.
  • Tannast M Department of Orthopaedic and Trauma Surgery, University of Bern, Inselspital, Freiburgstrasse 3, 3010, Bern, Switzerland.
  • Keel MJ Department of Orthopaedic and Trauma Surgery, University of Bern, Inselspital, Freiburgstrasse 3, 3010, Bern, Switzerland.
  • Bastian JD Department of Orthopaedic and Trauma Surgery, University of Bern, Inselspital, Freiburgstrasse 3, 3010, Bern, Switzerland. johannesbastian@gmx.de.
  • 2016-03-25
Published in:
  • Archives of orthopaedic and trauma surgery. - 2016
English Medial penetration of the helical blade into the hip joint after fixation of trochanteric fractures using the proximal femur nail antirotation (PFN-A) is a potential failure mode. In low demand patients a blade exchange with cement augmentation may be an option if conversion to total hip arthroplasty is unfeasible to salvage the cut-through. This article describes a technique to avoid intraarticular cement leakage using a cement plug to close the defect in the femoral head caused by the cut-through.