Journal article

Radiation dose in vertebroplasty.

Show more…
  • 2004-02-18
Published in:
  • Neuroradiology. - 2004
English We wished to measure the absorbed radiation dose during fluoroscopically controlled vertebroplasty and to assess the possibility of deterministic radiation effects to the operator. The dose was measured in 11 consecutive procedures using thermoluminescent ring dosimeters on the hand of the operator and electronic dosimeters inside and outside of the operator's lead apron. We found doses of 0.022-3.256 mGy outside and 0.01-0.47 mGy inside the lead apron. Doses on the hand were higher, 0.5-8.5 mGy. This preliminary study indicates greater exposure to the operator's hands than expected from traditional apron measurements.
Language
  • English
Open access status
green
Identifiers
Persistent URL
https://sonar.ch/global/documents/169157
Statistics

Document views: 29 File downloads:
  • Full-text: 0